GAMMA-THERAPY SYSTEM

SCo "Rawenstvo" is the only manufacturer of remote gamma-therapy systems in Russia. The company is among the global top five of unique developers of such kind of medical devices.

The field of application of gamma-therapy complex is irradiation of malignant tumors such as:

  • lung cancer
  • larynx cancer
  • breast cancer
  • bowel cancer
  • uterine corpus cancer
  • small pelvis lymph nodes
  • esophageal cancer, etc

1 Demountable gantry (for installation of the complex, technological openings in the walls are not required)

2 Radiation source with 9 800 Cu activity

3 Asymmetric collimator for fields from 2 × 2 cm to 35 × 35 cm with penumbras when using a trimmer less than 1 cm

4 Stop frame

5 Motorized table with four degrees of freedom

6 Radiolucent carbon deck

MAIN CONTROL PANEL

НOn the main control panel, it is possible to select one of two modes of operation: simulation (with the gate power circuit off) and procedure (irradiation plan execution mode). Access to the service mode with a key is provided.

MANUAL CONTROL PANEL

A manual control panel can be used for:

  • setting the treatment table to the first frame exposure by the DICOM protocol
  • setting the treatment table to the extreme position for patient placement
  • run simulations of sessions from the procedural
  • emergency stop work

SOFTWARE

Support for DICOM protocols. The complex supports the 3D planning system and has the ability to implement the "dynamic wedge" mode.

When the complex starts developed autotest implements self-diagnostics with a report on the readiness of the complex to work.

The complex keeps statistics of operation of all drives and sensors in automatic mode. The collected information can be broadcast to the manufacturer via the Internet or transmitted on a flash drive

Two-level control system of the complex: the control of the tripod drives is concentrated in the lower level controller, which is embodied on the modern element base in an industrial fault-tolerant computer.

Intuitive interface with the implemented schedule of patients irradiation, user support and service mode.

SOFTWARE

Support for DICOM protocols. The complex supports the 3D planning system and has the ability to implement the "dynamic wedge" mode.

When the complex starts developed autotest implements self-diagnostics with a report on the readiness of the complex to work.

The complex keeps statistics of operation of all drives and sensors in automatic mode. The collected information can be broadcast to the manufacturer via the Internet or transmitted on a flash drive.

Two-level control system of the complex: the control of the tripod drives is concentrated in the lower level controller, which is embodied on the modern element base in an industrial fault-tolerant computer.

Intuitive interface with the implemented schedule of patients irradiation, user support and service mode.

MAIN CHARACTERISTICS:

Distance from the source to focus distance (source-isocenter)
80 cm;
The height of the isocenter above the floor
130 cm;

GANTRY

Gantry has a modular design, consisting of three parts; can be delivered for installation in the treatment room through the maze.

Range of rotation of gantry
-220/+220;
Velocity of rotation gantry
from 0,5°/sec to 6°/sec;
Maximum activity of the Radiation source Co-60
9 800 Cu;

COLLIMATION SYSTEM

Demountable motorized trimmer to reduce penumbra (penumbra less than 1 cm)

Asymmetric collimator with rotation
-185/+185;
Motor-operated wedge
60°;
Free space from isocenter to collimator
35 cm;
Free space from isocenter to applicator
18 cm;
Removable applicator with the load of units (up to 20 kg)
Maximum field size in isocenter
35 cm × 35 cm;
Minimum field size in isocenter
2 × 2 cm;

TREATMENT TABLE

Table construction is isocentric with four degrees of freedom.

The range of vertical movement to the deck
from 62 to 180 cm;
Longitudinal movement to the deck
130 cm ;
The range of table rotation relative to isocenter
-95°/+95°;
Treatment table capacity
200 kg;